Haematocrit (HCT) is an important parameter for both adult and paediatric patients and can indicate potentially serious conditions. The most common assessment methods are microhaematocrit testing and automated analysers; however, developing countries have specific needs that these technologies do not always meet. Paper-based devices may be suitable in settings where tools need to be affordable, fast, easy to use and portable.
This study describes and validates against a reference method a new approach to estimating HCT from the penetration speed of a lateral-flow test strip, designed for use in low- and middle-income country (LMIC) settings.
To calibrate and test the method, researchers collected 145 blood samples from 105 healthy newborns born after 37 weeks of gestation: 29 calibration samples and 116 test samples, covering HCT values from 31.6% to 72.5%. A reflectance meter measured the time difference (Δt) from loading whole blood onto the test strip to saturation of the nitrocellulose membrane.
A non-linear relationship between HCT and Δt was observed and estimated with a third-degree polynomial equation (R² = 0.91), valid for HCT values from 30% to 70%. The model was then used to estimate HCT in the test set, showing good agreement with the reference method (r = 0.87, p < 0.001), a low mean difference of 0.53 ± 5.04%, and a slight tendency to overestimate higher HCT values. Mean absolute error was 4.29%, while maximum absolute error was 10.69%.
The proposed method is not sufficiently precise for diagnostic use, but may be suitable as a rapid, low-cost and easy-to-use screening tool, particularly in low- and middle-income settings.